工业CT旋转扫描台回转精度对成像质量的影响
The Influence of Rotation Table Rotating Accuracy on Reconstructed Image Quality of Industrial CT
-
摘要: 研究旋转扫描台回转精度与工业CT重建图像质量的关系, 对选择合适回转精度的旋转扫描台, 降低加工成本具有现实意义。通过计算机仿真分析, 模拟了旋转扫描台回转偏差分别服从正态分布、偏态分布的情况, 采用滤波反投影算法重建了不同回转精度情况下的工业CT图像, 并通过数值指标对工业CT重建图像的质量进行了定量评价。仿真结果表明, 旋转扫描台的绝对定位精度对工业CT重建图像质量影响较大, 而旋转扫描台的重复定位精度对工业CT重建图像质量影响较小。Abstract: The study on the influence of rotation table rotating accuracy on reconstructed industrial CT image quality was meaningful to select a suitable rotation table and reduce the processing cost of rotation table. In the simulation, the location errors of rotation table were divided into two types: normal distribution and skew distribution. And the projection data of different accuracy case were reconstructed by filtered back-projection algorithm. In addition, the quality of industrial CT image was evaluated by numerical indicators. The simulation results show that the absolute positioning accuracy of rotation table had a strong influence on reconstructed image quality, while the repeated positioning accuracy of rotation table had less effect on the image quality.