Advanced Search
    YAO Yan, ZHANG Xue-yi, ZOU Song-hua, LIU Hui-bin, QIN Tie-nan. Uncertainty Evaluation of Indication Error of Ultrasonic Coating Thickness Measurement[J]. Nondestructive Testing, 2017, 39(3): 10-12. DOI: 10.11973/wsjc201703003
    Citation: YAO Yan, ZHANG Xue-yi, ZOU Song-hua, LIU Hui-bin, QIN Tie-nan. Uncertainty Evaluation of Indication Error of Ultrasonic Coating Thickness Measurement[J]. Nondestructive Testing, 2017, 39(3): 10-12. DOI: 10.11973/wsjc201703003

    Uncertainty Evaluation of Indication Error of Ultrasonic Coating Thickness Measurement

    • Ultrasonic Thickness Measurement was widely used in processing and testing of parts for quality control. The uncertainty evaluation was significant in the indication Error. The uncertainty evaluation of indication error of ultrasonic thickness measurement was undertaken. The results show that the best estimate of Indication Error of the measurement was 19.8 μm and the uncertainty was 21.9 μm. Furthermore, the uncertainty of instrumental precision makes the greatest contribution to the uncertainty evaluation.
    • loading

    Catalog

      Turn off MathJax
      Article Contents

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return